4-4/5 digits 50,000
counts.
Selectable stable mode 5-4/5 digits 500,000
counts for DC voltage, & 6 digits 999,999
counts for Hz
POLARITY :
Automatic UPDATE RATE :
4-4/5 digits fast mode : 5 per second nominal;
5-4/5 digits stable mode : 1.25 per second
nominal;
42 Segments Bar graph : 60 per second max.
LOW BATTERY : Below approx. 7V
o OPERATING TEMPERATURE : 0 degree C to 45 degree C RELATIVE HUMIDITY :
Maximum relative humidity 80% R.H. For
temperature upto 31degree C decreasing linearly to
50% relative humidity at 45 degree C POLLUTION DEGREE : 2
STORAGE TEMPERATURE : -20 degree C to 60 degree C ALTITUDE : Operating below 2000m TEMPERATURE COEFFICIENT :
nominal 0.1 x (specified accuracy) /
degree C @ (0 degree C -- 18 degree C or 28 degree C -- 40
degree C), or otherwise
specified. SENSING : AC, AC + DC True RMS POWER SUPPLY : Single Alkaline 9V battery. POWER CONSUMPTION : 5mA typical APO TIMING : Idle for 4.5 minutes APO CONSUMPTION : 20mA DIMENSION :
186(L) mm x 87(W) mm x 35.5(H) mm;
198(L) mm x 97(W) mm x 55(H) mm
with Holster WEIGHT : 390 gm, 500 gm with Holster
SAFETY : The
meter is protected, against the
users, by double insulation per EN61010-1 and
IEC61010-1 2nd Edition (2001) to CAT III 1000V
& CAT IV 600V. The meter also meets UL3111-1
(1994) * and CSA C22.2 No. 1010-1-92* to
CAT III 1000V
*Category IV safety standard was first released
in IEC61010-1 2nd Edition in year 2001, and
was yet an UL published standard at the time
this catalog was written. E.M.C. : Meets EN61326(1997, 1998/A1),
EN61000-4-2(1995) and EN61000-4-3(1996)
OVERLOAD PROTECTIONS :
mA & mA: 1A/1kV(or 0.44A/1kV)IR 10kA,Ffuse
A :15A/1kV,IR 10kA(or 11A/1kV,IR 20kA),F fuse
V : 1050 Vrms, 1450 V peak
mV, W & Others : 1050 Vrms, 1450 V peak TRANSIENT PROTECTION :
8kV lightning surge (1.2/50ms)
STANDARD ACCESSORIES : Test leads (pair),
holster, Battery installed, User’s manual,
K-type thermocouple x 1. OPTIONAL ACCESSORIES : PC interface Kit
(RS232 optical adapter cable + RS232 Software
CD) , pins to type-K socket plug adapter. SPECIAL FEATURES : Record Max, MIN,
MAX-MIN readings. Crest (Instantaneous Pead
Hold) MAX, MIN, MAX-MIN readings. Relative
zero mode. 500,000 counts high resolution
stable reading mode. Backlighted display.
dBm readings. T1-T2 differential temperature
readings. %4-20 mA loop current readings. High
noise rejection filtered Line Level Frequency mode.
Data Hold. Audible & Visible input warning.